- Recalling firm
- Vesco Medical Llc
- Reason for recall
- The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the prescribed dose, if proper controls are not in place. Over- or under-delivery of enteral nutrition through the pump and associated feed sets could cause temporary or medically reversible adverse health consequences, with a remote probability of serious adverse health consequences.
- Affected products (lot / model / serial)
- UDI-DI 00850002550481; Lot Numbers: all lots manufactured through 08/12/2026
- Units affected
- 6,060 units
- Distribution pattern
- US Nationwide distribution.
- Firm location
- Westerville, OH, United States
- Recall type
- Voluntary: Firm initiated
- How the firm notified customers
- Letter
- Product type
- Devices
- Status
- Ongoing
- Recall number
- Z-0377-2027
- Recall initiated
- October 8, 2026
- FDA classification date
- November 6, 2026
- FDA report date
- November 18, 2026
- FDA event ID
- 100054
- Last updated on HTM Vibe